Technical Data
Axio Scope.A1
28
M60-2-0007 e 05/08
Transformer HBO 100 W
Range of application........................................................................................................................ indoor
Protection class......................................................................................................................................... I
Protection type .................................................................................................................................. IP 20
Voltage ................................................................................................................... 100 VAC ... 240 VAC
Power frequency ..................................................................................................................... 50 ... 60 Hz
Power input with HBO 103 in operation ........................................................................................ 155 VA
Fuses according to IEC 127
Microscope stand Axio Scope.A1 for LED illumination in transmitted light ........... 2x T 3.15 A/H, 5x20 mm
Microscope stand Axio Scope.A1 for HAL 50 illumination in transmitted light ...... 2x T 3.15 A/H, 5x20 mm
Transformer mbq52ac-z for HBO 50............................................................. ..........100 V, 127 V: 2x T 4 A
..........................................................................................................................220 V - 240 V: 2x T 2.5 A
Transformer HBO 100 W ............................................................................................T 2.0 A/H, 5x20 mm
Auxiliary power supply unit 12 V DC 100 W .......................................................... 2x T 5.0 A/H, 5x20 mm
Light sources
LED illumination DL
Power input ....................................................................................................................................7 W
Halogen lamp ........................................................................................................................ 12 V / 50 W
Light source control ................................................................infinitely variable from approx. 3 to 12 V
Halogen lamp ...................................................................................................................... 12 V / 100 W
Light source control ................................................................infinitely variable from approx. 3 to 12 V
Mercury vapor short arc lamp ........................................................................................................ HBO 50
Power input for HBO 50................................................................................................................50 W
Mercury vapor short arc lamp ...............................................................................................HBO 103 W/2
Power input for HBO 103 W/2 ....................................................................................................100 W
Illumination system Colibri
Power input ..................................................................................................................................70 W
Axio Scope.A1:
Stand with manual stage focusing
Rough drive .................................................................................................... approx. 4 mm / rotation
Fine drive ......................................................... approx. 0.4 mm / rotation; approx. 4 µm mark distance
Lift range ...................................................................................................................... approx. 25 mm
Vertical stop..........................................................................................................mechanically variable
Condenser 0.9/1.25 H with optional
Modulator disk ........................................................................................for bright field, dark field and
...........................................................................................................phase-contrast 1, 2, 3 or PlasDIC
Manual objective change ..............................................................................via nosepiece, 6-fold H, M27
Manual reflector module change ........................................................................via reflector slider 2-fold,
.................................................................................................................. reflector turret 4-fold or 6-fold
Содержание Axio Scope.A1
Страница 1: ...Operating Manual Axio Scope A1 Microscope for Routine and Entry Level Research...
Страница 22: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 22 M60 2 0007 e 05 08 2 2 Systems Overview...
Страница 23: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 23...
Страница 24: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 24 M60 2 0007 e 05 08...
Страница 25: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 25...
Страница 26: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 26 M60 2 0007 e 05 08...
Страница 122: ......
Страница 123: ...Click below to find more Mipaper at www lcis com tw Mipaper at www lcis com tw...