OPERATION
Carl Zeiss
Illumination and Contrasting Method
Axio Scope.A1
70
M60-2-0007 e 05/08
•
Center the field diaphragm image with both
centering screws (
620H620H
Fig. 4-1/
4
) on the condenser
carrier (
621H621H
Fig. 4-1/
C
). Open the field diaphragm
enough to make the edge of the diaphragm
disappear from the field of view (
622H622H
Fig. 4-1/
D
).
•
In order to adjust the aperture (contrast), take
an eyepiece from the tube barrel. Looking
through the barrel adjust the aperture with the
adjusting lever (
623H623H
Fig. 4-2/
4
) to the size of approx.
2/3 ... 4/5 of the diameter of the objective exit
pupil (
624H624H
Fig. 4-1/
E
). In most cases this aperture
gives the best contrast at almost full resolution
and is thus the best compromise for the human
eye.
•
Replace eyepiece in tube barrel.
Each objective changes the size of the object field and the objective aperture. The centering
might also vary slightly from objective to objective, so that an adjustment of field diaphragm
and aperture might become necessary for optimal results.
When using objectives < 10x, the front lens of the condenser (if swivel-mounted) must be
folded out and the aperture opened fully. In order to achieve a better contrast, the field
diaphragm can be pulled closer when working with large object fields. To do this you must
reduce the opening by a certain range. Avoid closing it too much in order to guarantee a
consistent illumination of the field of view.
A bright-field adjustment according to KÖHLER is not possible when using transmitted light
LED.
Fig. 4-1
Microscope adjustment in
transmitted light/bright-field
Содержание Axio Scope.A1
Страница 1: ...Operating Manual Axio Scope A1 Microscope for Routine and Entry Level Research...
Страница 22: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 22 M60 2 0007 e 05 08 2 2 Systems Overview...
Страница 23: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 23...
Страница 24: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 24 M60 2 0007 e 05 08...
Страница 25: ...TECHNICAL MANUAL Axio Scope A1 Systems Overview Carl Zeiss M60 2 0007 e 05 08 25...
Страница 26: ...TECHNICAL MANUAL Carl Zeiss Systems Overview Axio Scope A1 26 M60 2 0007 e 05 08...
Страница 122: ......
Страница 123: ...Click below to find more Mipaper at www lcis com tw Mipaper at www lcis com tw...