Appendix
App-1
IM WT310E-17EN
1
2
3
4
5
6
7
8
9
App
Index
Appendix 1
Error Messages
This section explains communication error messages.
• Error messages that are read from a PC such as through the :STATus:ERRor? command are
output in English.
• If servicing is necessary to solve the problem indicated by a message, contact your nearest
YOKOGAWA dealer.
• Only communication error messages are listed here. For details on other error messages, see
the Getting Started Guide, IM WT310E-02EN.
• Communication syntax errors
• Communication execution errors
• Device-specific and other errors
• Communication query errors
• System errors (communication)
• Information (1 to 99)
• Execution Errors (600 to 899)
• System Errors
100 to 199
200 to 299
300 to 399
400 to 499
300, 399
1 to 99
600 to 899
900 to 999
Listed below
Listed in section 6.2 of the
Getting Started Guide, IM WT310E-02EN
Communication Syntax Errors (100 to 199)
Error in communication command
Code
Message
Corrective Action
Page
102
Syntax error.
A syntax error not covered by error codes 100 to 199.
Chapters 5
and 6
103
Invalid separator.
Separate data values with a comma.
5-1
104
Data type error.
See page 5-6 and 5-7 and use the correct data type for
each parameter.
5-6 and
5-7
108
Parameter not allowed.
Check the number of data values.
5-6 and
chapter 6
109
Missing parameter.
Be sure to include all necessary data values.
5-6 and
chapter 6
111
Header separator error.
Use a comma to separate each header from its data.
5-1
112
Program mnemonic too long.
Check the command length.
Chapter 6
113
Undefined header.
Check the header.
Chapter 6
114
Header suffix out of range.
Check the header.
Chapter 6
120
Numeric data error.
A value must be specified where the syntax contains
<NRf>.
5-6
123
Exponent too large.
Where the syntax contains <NR3>, make the exponent
that follows E smaller.
5-6 and
Chapter 6
124
Too many digits.
Limit numeric values to 255 digits or less.
5-6 and
Chapter 6
128
Numeric data not allowed.
Use a data type other than <NRf>.
5-6 and
Chapter 6
131
Invalid suffix.
Check the unit of <Voltage>, <Current>, <Time>, or
<Frequency>.
5-6
134
Suffix too long.
Check the unit of <Voltage>, <Current>, <Time>, or
<Frequency>.
5-6
138
Suffix not allowed.
Only the following units can be used: <Voltage>,
<Current>, <Time>, <Frequency>.
5-6
141
Invalid character data.
Be sure to select one of the listed choices when the
syntax contains {...|...|...}.
Chapters 5
and 6
144
Character data too long.
Check the spelling of the strings when the syntax
contains {...|...|...}.
Chapter 6
148
Character data not allowed.
Use a data type other than {...|...|...}.
Chapter 6
150
String data error.
Enclose parameters with single or double quotation
marks where the syntax contains <String>.
5-7
Appendix