Appendix
App-3
IM WT310E-17EN
1
2
3
4
5
6
7
8
9
App
Index
Communication Query Errors (400 to 499)
Error in communication Query
Code
Message
Corrective Action
Page
410
Query INTERRUPTED.
Check the transmission and reception order.
5-2
420
Query UNTERMINATED.
Check the transmission and reception order.
5-2
430
Query DEADLOCKED.
Keep program messages to 1024 bytes or less in length,
including <PMT>.
5-2
440
Query UNTERMINATED after indefinite response. Do not write a query after *IDN? or *OPT?.
—
System Communication Errors (300 and 399)
Error in System Operation
Code
Message
Corrective Action
Page
300
Communication device-specific error.
Servicing is required.
—
399
Fatal error in the communication driver.
Servicing is required.
—
Communication Warning (50)
Warning
Code
Message
Corrective Action
Page
50
*OPC/? exists in message.
Write *OPC or *OPC? at the end of program messages.
—
Other Errors (350, 390)
Code
Message
Corrective Action
Page
350
Queue overflow.
Read the error queue.
5-6
390
Overrun error (RS-232 only)
Decrease the baud rate.
3-8
Note
Code 350 occurs when the error queue overflows. This error is only returned in response to a :STATus:
ERRor? query; it is never displayed on the screen.
Appendix 1 Error Messages