6-16
IM WT310E-17EN
6.8
INTEGrate Group
The commands in this group deal with integration.
You can make the same settings and queries that you can make by pressing INTEG SET, START, STOP, and RESET
on the front panel.
:INTEGrate:STATe?
Function
Queries the integration status.
Syntax
:INTEGrate:STATe?
Example
:INTEGRATE:STATE? -> RESET
Description • The response is as follows:
RESet = Integration reset
STARt = Integration in progress
STOP = Integration stop
ERRor = Abnormal integration termination
(integration overflow, power failure)
TIMeup = Integration stop due to integration
timeout
:INTEGrate?
Function
Queries all
integration settings.
Syntax
:INTEGrate?
:INTEGrate:MODE
Function
Sets or queries the
integration mode.
Syntax
:INTEGrate:MODE {NORMal|CONTinuous}
:INTEGrate:MODE?
NORMal
= Standard integration mode
CONTinuous
= Continuous integration mode
Example
:INTEGRATE:MODE NORMAL
:INTEGRATE:MODE?
-> :INTEGRATE:MODE NORMAL
:INTEGrate:TIMer
Function
Sets or queries the
integration timer value.
Syntax
:INTEGrate:TIMer {<NRf>,<NRf>,<NRf>}
:INTEGrate:TIMer?
{<NRf>,<NRf>,<NRf>}
=
0,0,0
to
10000,0,0
First
<NRf>
=
0
to
10000
(hours)
Second
<NRf>
=
0
to
59
(minutes)
Third
<NRf>
=
0
to
59
(seconds)
Example
:INTEGRATE:TIMER 1,0,0
:INTEGRATE:TIMER?
-> :INTEGRATE:TIMER 1,0,0
:INTEGrate:STARt
Function
Starts
integration.
Syntax
:INTEGrate:STARt
Example
:INTEGRATE:START
:INTEGrate:STOP
Function
Stops
integration.
Syntax
:INTEGrate:STOP
Example
:INTEGRATE:STOP
:INTEGrate:RESet
Function
Resets the integrated value.
Syntax
:INTEGrate:RESet
Example
:INTEGRATE:RESET