6-6
IM WT310E-17EN
:AOUTput[:NORMal]:RATE<x>
Function
Sets or queries the maximum and minimum
values for when the D/A output is in manual
range mode. Sets or queries the comparison
upper and lower limits for when the D/A output is
in comparator mode.
Syntax
:AOUTput[:NORMal]:RATE<x>
{<NRf>,<NRf>}
:AOUTput[:NORMal]:RATE<x>?
<x>
=
1
to
12
(output channel)
<Element>
=
{<NRf>|SIGMA}
<NRf>
=
-9.999E+12
to
9.999E+12
Example
:AOUTPUT:NORMAL:RATE1 100,-100
:AOUTPUT:NORMAL:RATE1?
-> :AOUTPUT:NORMAL:RATE1 100.0E+00,
-100.0E+00
Description • When the D/A output is in manual range mode
(:AOUTput[:NORMal]:MODE<x> MANual)
Set the rated value for +5 V output and then
that for –5 V output.
• When the D/A output is in comparator mode
(:AOUTput[:NORMal]:MODE<x> COMPare)
Set the upper limit and then the lower limit.
• When the D/A output is in fixed range mode
(:AOUTput[:NORMal]:MODE<x> FIXed)
There is no need to set these values. (The
values do not affect the output operation.)
6.2 AOUTput Group