6-5
IM WT310E-17EN
Commands
1
2
3
4
5
6
7
8
9
App
Index
:AOUTput?
Function
Queries all D/A output settings.
Syntax
:AOUTput?
:AOUTput[:NORMal]:PRESet
Function
Sets the D/A output items to their default values.
Syntax
:AOUTput[:NORMal]:PRESet
{NORMal|INTEGrate}
Example
:AOUTPUT:NORMAL:PRESET NORMAL
:AOUTput[:NORMal]:CHANnel<x>
Function
Sets or queries a D/A output item (function or
element).
Syntax
:AOUTput[:NORMal]:CHANnel<x> {NONE|
<Function>[,<Element>]}
:AOUTput[:NORMal]:CHANnel<x>?
<x>
=
1
to
12
(output channel)
NONE
= no output item
<Function>
=
{U|I|P|S|Q|LAMBda|
PHI|FU|FI|WH|WHP|WHM|AH|AHP|
AHM|MATH|UPeak|IPeak}
<Element>
=
{<NRf>|SIGMa}
(<NRf>
=
1
to
3)
Example
:AOUTPUT:NORMAL:CHANNEL1 U,1
:AOUTPUT:NORMAL:CHANNEL1?
-> :AOUTPUT:NORMAL:CHANNEL1 U,1
Description • For details on <Function>, see “Function
Option List (1)” on page 6-24.
• If <Element> is omitted, the element is set to 1.
• For {MATH}, <Element> does not need to be
specified. In responses, <Element> is omitted.
6.2
AOUTput Group
The commands in this group deal with D/A output.
You can make the same settings and queries that you can by pressing the UTILITY key on the front panel and then
using the dA menu or by pressing the INTEG SET key and then using the dAtimE menu. The commands in this group
are only valid on models with the D/A output (/DA4 or /DA12) option.
:AOUTput[:NORMal]:IRTime
Function
Sets or queries the rated integration time that is
used in the D/A output of the integrated value.
Syntax
:AOUTput[:NORMal]:IRTime
{<NRf>,<NRf>,<NRf>}
:AOUTput[:NORMal]:IRTime?
{<NRf>,<NRf>,<NRf>}
=
0,0,0
to
10000,0,0
First
<NRf>
=
0
to
10000
(hour)
Second
<NRf>
=
0
to
59
(minute)
Third
<NRf>
=
0
to
59
(second)
Example
:AOUTPUT:NORMAL:IRTIME 1,0,0
:AOUTPUT:NORMAL:IRTIME?
-> :AOUTPUT:IRTIME 1,0,0
:AOUTput[:NORMal]:MODE<x>
Function
Sets or queries a D/A range mode.
Syntax
:AOUTput[:NORMal]:MODE<x> {FIXed|
MANual|COMPare}
:AOUTput[:NORMal]:MODE<x>?
<x>
=
1
to
12
(output channel)
Example
:AOUTPUT:NORMAL:MODE1 FIXED
:AOUTPUT:NORMAL:MODE1?
-> :AOUTPUT:NORMAL:MODE1 FIXED
Description • FIXed = Fixed range mode (default value)
O5 V when the rated value of each
measurement function is received.
• MANual = Manual range mode
The displayed values of the measurement
function when +5 V and –5 V are output as D/A
output can be set to any values of your choice.
This enables the D/A output to be expanded or
reduced for each channel (D/A zoom).
• COMPare = Comparator mode
By comparing with the comparator limits, this
instrument o5 V, 0 V, or –5 V. Relay
contact output, like that of the WT210/WT230,
can be emulated by driving a relay with the
output voltage.
Chapter 6
Commands