2-22
IM WT310E-01EN
Configuration When the Data Update Interval Is Auto
Set the following items.
Synchronization Source
• The synchronization source that you will set in this section is a special synchronization source that
is used only when the data update interval is Auto is used.*
* For details on synchronization sources, see section 2.8.
• You can select the synchronization source from the options listed below. The default value is U1 for
the WT310E/WT310EH and I1 for the WT332E/WT333E.
• U1: The synchronization source is set to the voltage of element 1.
• I1: The synchronization source is set to the current of element 1.
• U2: The synchronization source is set to the voltage of element 2. (WT333E only)
• I2: The synchronization source is set to the current of element 2. (WT333E only)
• U3: The synchronization source is set to the voltage of element 3. (WT332E and WT333E)
• I3: The synchronization source is set to the current of element 3. (WT332E and WT333E)
Timeout
• Timeout is the time limit for detecting the period of the input waveform.
• Select 1s, 5s, 10s, or 20s. The default value is 1s.
• If the input signal frequency is low and the period of the synchronization source cannot be detected
within the timeout period, the frequency data will be outside the measurement range and will result
in error. The measurement functions of normal measurement determine measured values using the
entire period up to the timeout as the measurement period.
Functional Limitations
The following limitations apply when the data update interval is set to Auto.
• The integration function cannot be used. If you try to start it, error 886 will occur.
• The measurement data storage function cannot be used. If you try to start it, error 886 will occur.
2.10 Setting the Data Update Interval