![YOKOGAWA WT310E Скачать руководство пользователя страница 74](http://html.mh-extra.com/html/yokogawa/wt310e/wt310e_user-manual_911416074.webp)
5-4
IM WT310E-01EN
Displayed Integrated Value Functions
You can display integrated values for different polarities separately by choosing the appropriate
functions to display.
Displayed Function
Measurement Mode
Display Details
Wh
RMS, VOLTAGE MEAN, DC
Sum of positive and negative watt hours
Wh±
1
RMS, VOLTAGE MEAN, DC
Positive watt hours
Wh±
1
RMS, VOLTAGE MEAN, DC
Negative watt hours
Ah
RMS, VOLTAGE MEAN
Sum of ampere hours
DC
Sum of positive and negative ampere hours
Ah±
2
RMS, VOLTAGE MEAN
Total ampere hours (same as Ah)
DC
Positive ampere hours
Ah±
2
RMS, VOLTAGE MEAN
–0 is displayed
DC
Negative ampere hours
1 When Wh is the displayed function, the displayed function will be Wh± whether you press the
FUNCTION key once or twice. When you press the function key once and switch to Wh±, the
positive watt hour will be displayed. If you press the FUNCTION key once more, the negative watt
hour for Wh± will be displayed. When the negative watt hour is being displayed, the displayed value
is preceded by a minus sign.
2 When Ah is the displayed function, the displayed function will be Ah± whether you press the
FUNCTION key once or twice. When you press the function key once and switch to Ah±, the
positive ampere hour will be displayed. If you press the FUNCTION key once more, the negative
ampere hour for Ah± will be displayed. When the negative watt hour is being displayed, the
displayed value is preceded by a minus sign.
Note
• When negative integrated values are displayed, the minimum display reading will become –99999 MWh/
MAh because of the added minus character.
• During the period from when integration is started until it is reset, there are limits to the changes that you
can make to other functions. For details, see section 5.4.
Using an External Signal to Control Integration (Option)
On models with the D/A output option (/DA4 or /DA12), you can use the remote control feature to use
an external signal to start, stop, and reset integration. For information about the remote control feature,
see section 5.2 in the Getting Started Guide, IM WT310E-02EN.
5.1 Integration Feature