2-8
IM WT310E-01EN
Checking a Range
To check the currently set range during measurement, press VOLTAGE or CURRENT. The set range
will appear on the display.
To return to the measured value display in menu configuration mode, press the same key again. In
quick configuration mode, the range is only displayed for the duration of the data update interval before
the measured value display is restored.
Measured value display
Measured value display
Power Range
The measurement range (power range) for measuring active power, apparent power, and reactive
power is determined as indicated below in accordance with the wiring system, voltage range, and
current range.
Wiring System
Power Range
1P2W (single-phase, two-wire)
Voltage range × current range
1P3W (single-phase, three-wire)
Voltage range × current range × 2
3P3W (three-phase, three-wire)
3V3A (three-voltage, three-current method)
3P4W (three-phase, four-wire)
Voltage range × current range × 3
• The maximum display is 99999 (when the number of displayed digits is set to 5).
• When the result of multiplying the voltage range by the current range reaches or exceeds 1000
W, the unit on the display will change to “kW.” When the result exceeds 1000 kW, the unit on the
display will change to “MW.”
• For specific voltage and current range combinations and power range values, see appendix 3.
Note
• In auto range mode, because the voltage and current ranges switch independently according to range
increase and decrease conditions, different power ranges may be set for the same measured or computed
power value.
• If you open the voltage input terminal, a voltage value of up to 0.3 V may be displayed due to hum noise
at the power line frequency of 50 Hz or 60 Hz and other phenomena. This is because of the high input
resistance of the voltage input terminal. Shorting the terminal will result in a value of 0 V.
2.3 Setting the Measurement Range When Using Direct Input