Standard Integration Mode
You can set a relative integration time (set a timer). Integration starts when you press START. When
one of the conditions below is met, integration is stopped, and the integration time and integrated value
are held.
• The specified timer time elapses.
• You press STOP.
• The integrated value reaches its maximum or minimum displayable value.
Reset
Start
Hold
Hold
Timer value
Integration
time
Integrated
value
Continuous Integration Mode
You can set a relative integration time. Integration starts when you press START. When the specified
integration timer time elapses, integration is automatically reset and restarted. Integration repeats until
you press Stop. If either of the conditions below is met, integration is stopped, and the integration time
and integrated value are held.
• You press STOP.
• The integrated value reaches its maximum or minimum displayable value.
Start
Reset
Hold
Hold
Timer
value
Timer
value
Timer
value
Stop
Integrated
value
Integration
time
Integration Methods (Equations)
For information on the equations used for integration, see appendix 1.
1.5 Integrated Power (Watt hour)