13-18
IM WT3001E-51EN
• Display Updating during the Cycle-by-Cycle Measurement
Even if you start the cycle-by-cycle measurement, [-------] (no data) is displayed until
the measurement is completed. When the measurement is completed, the measured
data is displayed.
• Switching the Element for Displaying the Measurement Data
• You can select the element/wiring unit from the choices below. The selectable items
vary depending on the installed elements.
Element1, Element2, Element3, Element4, ΣA, and ΣB
• If there are no elements that are assigned to the selected wiring unit, there is
no numeric data. Thus, [-------] (no data) is displayed in this case. For example,
if elements are assigned to ΣA and no elements are assigned to ΣB, then the
measurement function for ΣB shows [-------] (no data).
• Page Scrolling the Screen
A single page displays a list of measured data of 20 measurement cycles. If the
measurement cycle count exceeds 20, the data is displayed on multiple pages. In this
case, you can scroll or jump the page to switch the displayed page.
• Resetting the Cycle-by-Cycle Measurement
To restart the cycle-by-cycle measurement, execute reset and restart the
measurement when the cycle-by-cycle measurement status is indicating Complete.
You start the measurement in the Complete status.
Reset the cycle-by-cycle measurement to change the following settings.
• Measurement conditions (chapter 4 in the
User’s Manual IM WT3001E-01EN
)
• Measurement conditions of the cycle-by-cycle measurement (section 13.3)
• Flow Chart of the Cycle-by-Cycle Measurement
(Transition Diagram of the Cycle-by-Cycle Measurement Status)
Abort measurement [Reset]
(see section 13.4)
Re-measure [Reset]
(see section 13.4)
Reset
Change the measurement
conditions (see section 13.3)
Start measurement [Start] (see section 13.4)
Time Out
Measurement complete
Start
Complete
Time Out
Freq Err
Reference sections are indicated in parentheses.
Soft keys that you operate are indicated in brackets.
Note
You can save the measured data of the cycle-by-cycle measurement to a file in CSV format.
For the procedure to save the data, see section 9.3 in the
User’s Manual IM WT3001E-01EN
.
To save the data, select Numeric for the item to be saved.
13.4 Executing the Cycle-by-Cycle Measurement