1-10
IM WT3001E-51EN
1.5 Selecting the Line Filter and Synchronization
Source
Procedure
START STOP
REMOTE
CAL
MOTOR
SET
CURSOR
NULL
MENU
STORE SET
MENU
LOWER ITEM
LOWER FORM
ALL
USER SET
DISPLAY
ITEM & ELEMENT
RESET
SET
PAGE
PAGE
SCALING
HRM SET
MEASURE
SYNC SOURCE
AVG
LINE FILTER
FREQ
FILTER
FILE
MISC
STORE
IMAGE SAVE
HELP
NUMERIC
WAVE
OTHERS
ITEM
FORM
U / I / P
WP/q/
TIME
USER
ELEMENT
UPDATE
RATE
HOLD
SINGLE
INTEG
LOCAL
SHIFT
Cursor keys
To exit the menu during operation, press
ESC
located above the
soft keys.
In the procedural explanation below, the phrase “press the
cursor
keys
” may be used. This phrase refers to the procedures for
selecting items and entering values and character strings. For
details on the procedures, see section 3.14 in the
User’s Manual,
IM WT3001E-01EN
.
The settings covered in this section apply to the case when the input signal type is set to
Analog. You do not have to specify the settings in this section if the input signal type is
set to Pulse.
1.
Press
SHIFT+SCALING (MOTOR SET)
to display the Motor Set dialog box.
• Selecting the Line Filter
2.
Press the
cursor keys
to select Line Filter.
3.
Press
SET
. A line filter selection box appears.
4.
Press the
cursor keys
to select a value between OFF and 50 kHz.
5.
Press
SET
to confirm the line filter. The same filter is set to both the revolution
signal and the torque signal.
• Selecting the Synchronization Source
2.
Press the
cursor keys
to select Sync Source.
3.
Press
SET
. A synchronization source selection box appears.
4.
Press the
cursor keys
to select any of the elements/wiring units starting with U1.
5.
Press
SET
to confirm the synchronization source.