8-8
IM WT3001E-51EN
Frequency Measurement
Item
Specifications
Measurement range
• PLL synchronization method: 2.5 Hz ≤ f ≤ 100 kHz
• External sampling clock method: 0.15 Hz ≤ f ≤ 5 kHz
Measurement Function
Item
Specifications
Display update
Depends on the PLL source
• PLL synchronization method: 1 s or more
• External sampling clock method: 20 s or more
PLL Lock Timeout Warning
Item
Specifications
Timeout value
Depends on the PLL source
• PLL synchronization method: 5 s or more
• External sampling clock method: 40 s or more
Functional Limitations
Item
Specifications
Display
Waveform display, trend display, computed waveform display, and FFT display are not allowed.
Averaging
Moving average is not allowed.
Data update interval
Depends on the PLL source
Voltage/Current mode
Not selectable. The LED is OFF.
Peak measurement
Not allowed
Efficiency computation
Not allowed
Compensation
Efficiency compensation is invalid.
Integration
Not allowed
Store
Integration synchronization (Integ Sync) is not allowed. Waveform storage is not allowed.
File save
Waveform and waveform sampling data (Acquisition) are not allowed
NULL
Not allowed
Synchronized measurement Not allowed
Motor evaluation function
Not allowed
Integration synchronization using auto print is not allowed.
Delta computation
Not allowed
Measurement Functions
For a description of the measurement functions of wide bandwidth harmonic
measurement mode and how they are determined, see section 7.11. Measurement is not
allowed for measurement functions related to the functions above that have limitations.
For a list of measurement functions that cannot be measured, see appendices 10 and 11
in the
User’s Manual IM WT3001E-01EN
.
8.5 Wide Bandwidth Harmonic Measurement Specifications