3-5
IM WT3001E-51EN
D/A
Output and Remote Control (Option)
1
2
3
4
5
6
7
8
9
10
11
12
13
Index
Setting the Rated Time of Integrated D/A Output
13.
Press
INTEG
to display the Integ menu.
14.
Press the
Integ Set
soft key to display the Integ Set menu.
15.
Press the
D/A Output Rated Time
soft key to display the D/A Rated Time dialog
box.
16.
Press the
cursor keys
to select the hour, minute, or second box.
17.
Press
SET
. An entry box appears.
18.
Press the
cursor keys
to set the hour, minute, or second that you selected in step 16.
19.
Press
SET
or
ESC
to close the box.
20.
Repeat steps 16 to 19 to set the hour, minute, and second.
Checking the D/A Output Items
1.
Press
SHIFT+WIRING(INPUT INFO.)
. Power Element Settings appears.
2.
Press
PAGE
to show DA Output Settings (a list of D/A output settings). The
relation table of D/A output channels and measurement functions is displayed.
3.2 Setting the D/A Output