1-5
IM WT3001E-51EN
Motor Evaluation Function (Option)
1
2
3
4
5
6
7
8
9
10
11
12
13
Index
• Selecting the Torque Signal Type
2.
Press the
cursor keys
to select the Sense Type under Torque.
3.
Press
SET.
A signal type selection box appears.
4.
Press the
cursor keys
to select Analog or Pulse.
5.
Press
SET
to confirm the signal type.
Explanation
• Selecting the Signal Type
The signal applied from the revolution sensor or torque meter to this instrument can
be selected from the following two types.
• Analog
Select this when the signal type is a DC voltage (analog signal).
• Pulse
Select this when the signal type is a pulse signal.
• Signal Type and Setup Items
Some of the settings described in the subsequent sections are not required depending
on the signal type as described below.
• Settings Related to the Revolution Signal Type
Revolution Signal Type
Analog
Pulse
Analog range (section 1.4)
Required
Not required
Line filter (section 1.5)
Required
Not required
Synchronization source (section 1.5)
Required
Recommended*
Pulse range (section 1.6)
Not required
Required
Pulse count per revolution (section 1.6)
Not required
Required
* Measurement is possible using None (default), but it is recommended that you specify the
setting to improve the measurement accuracy. For details, see the explanation in section
1.5.
• Settings Related to the Torque Signal Type
Torque Signal Type
Analog
Pulse
Analog range (section 1.4)
Required
Not required
Line filter (section 1.5)
Required
Not required
Synchronization source (section 1.5)
Required
Recommended*
Pulse range (section 1.6)
Not required
Required
Pulse rating (section 1.6)
Not required
Required
* Measurement is possible using None (default), but it is recommended that you specify the
setting to improve the measurement accuracy. For details, see the explanation in section
1.5.
1.3 Selecting the Type of the Revolution and Torque Signals