17-3
IM WT1801-01EN
• If you enable automatic CSV conversion (Auto CSV Conversion) and set the storage destination to a
USB memory device, the maximum storage count is computed with 20% of the available USB memory
designated as valid for stored data (*.WTS and *.HDS files).
• If the maximum storage count is displayed as 0, there is not enough available memory in the storage
destination. Make more space available, by deleting files for example.
• If you change the number of stored items specified in
after you set the storage count, the
maximum storage count will change, and you will need to set the storage count again.
Storage Interval (Interval)
You can set the interval at which numeric data is stored.
• You can set the hour, minute, and second within the following range. If you set the interval to 00:00:00, the
numeric data will be stored at the numeric data update interval.
00:00:00 to 99:59:59
• When the storage mode is set to Integ Sync, storage continues even when integration is reset by the
. When the integration timer is reset, the storage interval timer is also reset.
• This setting is invalid when the storage mode is set to Event or Single Shot.
Scheduled Storage Times for Real-Time Storage Mode (Real Time Control)
These settings are only valid when the storage mode is set to Real Time. You can set the year, month, day, hour,
minute, and second of the storage start and stop times. Be sure to set the storage stop time to a time after the
storage start time. You can set the values within the following ranges.
• Year:
Any four-digit Gregorian calendar value
• Hour:Minute:Second:
00:00:00 to 23:59:59
• Now:
The scheduled storage start time is set to the current time.
• Copy:
The scheduled storage start time is copied into the scheduled storage stop time.
• You can set the day for the scheduled start or stop time in February to a value as high as the 31st day. If
you do so, an error message will appear when you start storage. Reset the scheduled start and stop times.
• The WT1800 recognizes leap years when it executes the storage operation.
Trigger Event for Event-Synchronized Storage Mode (Trigger Event)
This setting is only valid when the storage mode is set to Event. You can select the user-defined event whose
occurrence will cause storage to start. If you select a disabled user-defined event, storage cannot start.
For information about configuring user-defined events, see “User-Defined Events.”
Storage of the Numeric Data at Store Start (Store At Start)
• Select whether to store the numeric data at store start.
• You can configure this setting when:
• The storage mode is set to Manual, and the storage interval is not 00:00:00.
• The storage mode is set to Real Time, and the storage interval is not 00:00:00.
• The storage mode is set to Integ Sync.
17 Data Storage