16-10
IM WT1801-01EN
High Speed Data Capturing Operations
Examples for when the trigger mode is Auto or Normal.
Time
Start key
Completion
of initialization
High speed
data capturing
state
(HS State)
Ready
Start
Ready
Stop key or capture count
Completion of
file recording
File recording
state
(File State)
Ready
Rec
Trigger
Ready
Data
capturing
Trigger
wait
After a trigger occurs, measured data is regularly written to
the high speed data capturing data file (.WTS).
16 High Speed Data Capturing