Capture Control Settings (Control Settings)
Voltage and Current Measurement Modes (U/I Measuring Mode)
Select the modes for measuring voltage and current.
Setting
Select the method for setting the voltage and current measurement modes from the following options.
• Each:
Set the mode separately for the voltage and current of each input element.
• All:
Set the mode for the voltages and currents of all installed input elements at the same time.
U1 to I6
Select the voltage and current measurement modes from the following options.
rms, mean, dc, r-mean
For details about how the voltage and current values are determined in each measurement mode, see appendix
1 in the getting started guide, IM WT1801-03EN.
The voltage or current measurement mode
(rms, mean, dc, r-mean)
If the voltage and current measurement mode settings differ for elements assigned to the same wiring unit,
the voltage (UΣ) and current (IΣ) of the wiring unit are not measured and are displayed as "-------" (no data).
HS Filter (HS Filter)
Turning on the HS (high speed) filter stabilizes the measured values. The HS filter averages the measured
values. The HS filter is different from the
Set whether to enable the HS filter, and set the cutoff frequency. You can set the cutoff frequency within the
following range.
1 Hz to 1000 Hz (in 1 Hz steps)
Selecting OFF disables the HS filter.
You can stabilize the measured voltage (U), current (I), and power (P) values for each element and the
measured UmeanΣ, ImeanΣ, Ur-meanΣ, and Ir-meanΣ values for each wiring unit by specifying a low cutoff
frequency for the HS filter. However, the response to changes in the measurement data becomes slower.