10-8
IM WT1801-01EN
Waveform Mapping (Wave Mapping)
Assignment Method (Mode)
You can assign waveforms to windows on the screen. Choose one of the following methods for assigning the
waveforms to the windows.
• Auto
The waveforms whose displays are turned on are assigned in order according to their element numbers, with
an element’s voltage waveform (U) coming first, followed by its current (I), speed, torque, Aux1, and then
Aux2 waveform.
1, 2
• Fixed
Regardless of whether their displays are on or off, waveforms are assigned in order according to their element
numbers, with an element’s voltage waveform (U) coming before its current waveform (I). The speed and
Aux1 waveforms appear in the top window.
1, 2
The torque and Aux2 waveforms appear in the second window
from the top.
1, 2
• User
You can pick which waveforms to assign to which windows, regardless of whether the waveform’s displays
are on or off. You can set the display position to a number from 0 to 5. Number 0 corresponds to the window
at the top of the screen, and the window number increases for each successively lower window.
1 Speed and torque waveforms are available on models with the motor evaluation option.
2 Aux1 and Aux2 waveforms are available on models with the auxiliary input option.
Display Item (ITEM)
Displaying All Waveforms (All ON)
The waveforms of all input signals are displayed.
Displaying No Waveforms (All OFF)
No waveforms are displayed.
Selecting Which Waveforms to Display (Display ON/OFF)
• You can select whether to display the waveform of each input signal of each element by selecting or clearing
the signal’s check box. Only the input signals of installed elements appear.
• On models with the motor evaluation option, you can turn the displays of the waveforms of the speed and
torque input signals on and off.
• On models with the auxiliary input option, you can turn the displays of the waveforms of the Aux1 and Aux2
input signal on and off.
Vertical Zoom (Vertical Zoom)
Each displayed waveform can be scaled. You can select one of the following zoom factors:
0.1, 0.2, 0.25, 0.4, 0.5, 0.75, 0.8, 1, 1.14, 1.25, 1.33, 1.41, 1.5, 1.6, 1.77, 2, 2.28, 2.66, 2.83, 3.2, 3.54, 4, 5, 8,
10, 12.5, 16, 20, 25, 40, 50, or 100
10 Waveform Display