<4. Setting converter>
4-8
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
4.2
Contact input settings
Contact input settings > Contact input Wash, or Contact input change range.
When mA Output/Input is “-A2”, only Contact input 1 is available. When mA Output/Input is “-A4”,
both Contact input 1 and 2 (Ad) are available. When mA Output/Input is “-A2”, configuration of
Contact input 2 (Ad) does not affect the operation.
4.2.1
Contact input Type
Define one contact input from “Disabled”, “Contact input Wash”, “Contact input change range”. If
none of these is needed, select “Disabled”, otherwise setting error might occur.
4.2.2
Contact input Wash
Contact input settings > Contact input Wash > S1/S2/S3
Select one contact output out of S1/S2/S3 to have a contact input wash.
While doing so, go back to the contact output settings. Set “Wash” for S1/S2/S3. If S1/S2/S3 you
selected is not “Wash”, an error occurs. For detail information on contact output settings, read
4.4.
4.2.3
Contact input change range
Contact input settings > Contact input change range > mA1/mA2/mA3/mA4
On this screen, you can select the range of mA output from mA1, mA2, mA3 and mA4. For detail
information on settings of contact input change range (DI change range) read 4.3.1.