5.1 Generation of Standard Waveform and Numerical Expression Waveform
recommended to use strings different from those in “Table 5-2 Built-in Constants”.
When the
All Page OK
button is clicked, the constant takes effect on the page where the constant
is defined and subsequent pages.
For example, the constant defined on the first page affects all the pages. The constant defined on the
third page affects the third and subsequent pages, but not the first and second pages.
When the waveform on the displayed page is created with the
Page OK
button, the constant affects
the displayed page only. For example, the constant defined on the second page does not affect the
first and third pages.
To use the constant on other pages, click the
Compute
button to execute calculation on the page
where the constant is defined in advance.
b) Built-in constants
The constants shown in “Figure 5-2 Waveform Generation Screen - Waveform Function” were
included into
ARB Edit
at factory.
If the constant defined in “a) constant” has the same name as the built-in constant, the defined value
or expression takes effect.
Table 5-2 Built-in Constants
Name
Meaning
Value
pi
Circle ratio
3.1415926535898
c
Light speed
2.99792458e8
h
Planck constant
6.6260755e-34
k
Boltzmann constant
1.380658e-23
r
Euler constant
0.57721566490153
c) Functions
Specify a numerical expression in the Y=f (X) format in the numerical expression input section
[Y=]
.
Here, “X” is a value that varies within the range of the page. For example, when the page range is
1000 to 2000,
[X]
in the expression changes between 1000 and 2000.
The
[X]
value in the numerical expression is affected by the horizontal axis unit of the system setting
screen settings.
“5.6.2 Axis unit setting”
For example, assume that expressions such as
[Y=]
or
[sin (x)]
are used to specify a sine wave. The
argument for the sin() function is a radian expression.
Assume that the horizontal axis unit is
[Address]
and the range is between 0 and 8,192. Because
8,192 = 2*
π
*1303.797..., the result of this expression is a sine wave with approximately 1,304
waves.
IM FG410-61EN
5-6
Содержание FG410
Страница 1: ...User s Manual FG410 FG420 ARBITRARY WAVEFORM EDITOR IM FG410 61EN 3rd Edition ...
Страница 3: ......
Страница 10: ...1 OVERVIEW 1 1 General 1 2 1 2 Function Tree 1 2 1 3 Conventions 1 5 ARB Edit 1 1 ...
Страница 15: ...1 3 Conventions IM FG410 61EN 1 6 ...
Страница 44: ...4 7 Copy and Paste IM FG410 61EN 4 15 ...
Страница 45: ......
Страница 76: ...6 ERROR MESSAGES 6 1 Arbitrary Wave Edit Software Error 6 2 IM FG410 61EN 6 1 ...
Страница 80: ......
Страница 82: ......
Страница 83: ...https www yokogawa com ymi Yokogawa Test Measurement Corporation ...