<2. Specifications>
2-4
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
Functions (inclused Humidity Analyzer)
Display Function:
Displays values of the measured oxygen concentration, moisture
quantity, mixture ratio, etc.
Alarm, Error Display: Displays alarms such as “AL-06” or errors such as “Err-01” when any
such status occurs.
Calibration Functions:
Automatic calibration; Requires the Automatic Calibration Unit. It calibrates
automatically at specified intervals.
Semi-auto Calibration; Requires the Automatic Calibration Unit. Input calibration start
signal by optical switch or contact, then it calibrates automatically
afterwards.
Manual Calibration;
Calibration with opening/closing the valve of calibration gas in
operation interactively with the optical switch.
Maintenance Functions:
Can operate updated data settings in daily operation and
checking. Display data settings, calibration data settings, test
settings (current output loop check, input/output contact check).
Setup Functions:
Initial settings suit for the plant conditions when installing the converter.
Current output data settings, alarm data settings, contact data settings,
other settings.
Display and setting content:
Display Related Items: Oxygen concentration (vol% O
2
), output current value (mA), air ratio,
moisture quantity (in hot gases) (vol% H
2
O), mixture ratio(kg/kg), relative
humidity(%RH), dew point (°C), Cell temperature (°C ), thermocouple
reference junction temperature (°C ), maximum/minimum/average
oxygen concentration (vol% O
2
), maximum/ minimum/average moisture
quantity (vol% H
2
O), maximum/minimum/average mixture ratio (kg/
kg), cell e.m.f. (mV), cell internal resistance (Ω), cell condition (in four
grades), heater on-time rate (%), calibration record (ten times), time
(year/month/day/hour/minute), output 1, 2 current (mA), cell response
time (seconds),
Calibration Setting Items: Span gas concentration (vol% O
2
), zero gas concentration (vol%O
2
),
calibration mode (auto, semi-auto, manual), calibration type and method
(zero-span calibration, zero calibration only, span calibration only),
stabilization time (min.sec), calibration time (min.sec), calibration interval
(day/hour), starting time (year/month/day/hour/minute)
Output Related Items: Analog output/output mode selection, output conditions when warming-
up/maintenance/calibrating/abnormal, 4 mA/20 mA point oxygen
concentration (vol% O
2
), time constant, preset values when warming-up/
maintenance/calibrating/abnormal, output preset values on abnormal
Alarm Related Items: Oxygen concentration high alarm/high-high alarm limit values (vol% O
2
),
oxygen concentration low alarm/low-low alarm limit values (vol% O
2
),
oxygen concentration alarm hysteresis (vol% O
2
), moisture quantity high
alarm/ high-high alarm limit values (vol% H
2
O), moisture quantity low
alarm/ low-low alarm limit values (vol% H
2
O), mixture ratio high alarm/
high-high alarm limit values (kg/kg), mixture ratio low alarm/ low-low
alarm limit values (kg/kg), moisture quantity alarm hysteresis (vol%
H
2
O), mixture ratio alarm hysteresis (kg/kg), oxygen concentration/
moisture quantity/mixture ratio alarm detection,alarm delay (seconds)
Contact Related Items: Selection of contact input 1 and 2, selection of contact output 1 and
2 (abnormal, high-high alarm, high alarm, low alarm, low-low alarm,
maintenance, calibrating, range switching, warming-up, calibration gas
pressure decrease, flameout gas detection (answer-back of contact
input)