<2. Specifications>
2-3
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
Sample Gas Temperature:
0 to 700°C
It is necessary to mount the cell using inconel cell-bolts when the
temperature is greater than 600°C.
High temperature service ― greater than 700°C ― is not
available.
Sample Gas Pressure: -5 to +250 kPa (When the pressure in the furnace exceeds 3 kPa, it is
recommended to use pressure compensated type. When the pressure in
the furnace exceeds 5 kPa, pressure compensated type is required.)
No pressure fluctuation in the furnace should be allowed.
Note: When the detector is used in conjunction with a check valve and the ZA8F Flow Setting Unit, the maximum pressure of sample
gas is 150 kPa. When with a check valve and the ZR20H Automatic Calibration Unit, it is 200 kPa. If the pressure of your sample
gas exceeds these limits, consult with Yokogawa.
Probe Length:
0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0 m
Probe Material:
SUS 316 (JIS)
Ambient Temperature: -20 to +55°C (- 5 to +70°C on the case surface)
Storage Temperature: -30 to +70°C
Humidity Range:
0 to 95%RH (non-condensing)
Power Supply Voltage: Ratings; 100 to 240 V AC
Acceptable range; 85 to 264 V AC
Power Supply Frequency:
Ratings; 50/60 Hz
Acceptable range; 45 to 66 Hz
Power Consumption: Max. 300 W, approx. 100 W for ordinary use.
Reference Gas System:
Natural Convection, Instrument Air, or Pressure Compensated
Instrument Air System (excluding Natural Convection):
Pressure;
200 kPa + the pressure inside the furnace (It is recommended to use air
which is dehumidified by cooling to dew point -20°C or less, and dust or
oil mist are removed.)
Consumption; Approx. 1Nl/min
Wetted Material:
SUS 316 (JIS), Zirconia, SUS304 (JIS) or ASTM grade 304 (flange),
Hastelloy B, (Inconel 600, 601)
Construction: Heater and thermocouple replaceable construction. Non explosion-proof JIS
C0920 / equivalent to IP44D. Equivalent to NEMA 4X/IP66 (Achieved when the
cable entry is completely sealed with a cable gland in the recirculation pressure
compensated version.)
Gas Connection:
Rc1/4 or 1/4NPT(F)
Wiring Connection:
G1/2, Pg13.5, M20 x 1.5mm, 1/2NPT select one type (4 pieces)
Installation:
Flange mounting
Probe Mounting Angle: Horizontal to vertically downward.
When the probe insertion length is 2 m or less, installing at angles from
horizontal to vertically downward is available.
When the probe insertion length is 2.5m or more, mount vertically
downward (within ± 5°), and if installing at angles from horizontal to
vertically downward (within ± 5°), use a probe protector.
Case:
Aluminum alloy
Paint Color: Cover; Mint green (Munsell 5.6BG3.3/2.9)
Case; Mint green (Munsell 5.6BG3.3/2.9)
Finish:
Polyurethane corrosion-resistance coating
Weight:
Insertion length of 0.4m: approx. 8 kg (JIS 5K 65) / approx. 13 kg (ANSI 150 4)
Insertion length of 1.0m: approx. 10 kg (JIS 5K 65) / approx. 15 kg (ANSI 150 4)
Insertion length of 1.5m: approx. 12 kg (JIS 5K 65) / approx. 17 kg (ANSI 150 4)
Insertion length of 2.0m: approx. 14 kg (JIS 5K 65) / approx. 19 kg (ANSI 150 4)
Insertion length of 3.0m: approx. 17 kg (JIS 5K 65) / approx. 22 kg (ANSI 150 4)