<12. Troubleshooting>
12-4
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
12.2 Displays and Measures to Take When
Alarms are Generated
12.2.1
What is an Alarm?
When an alarm is generated, the alarm indication blinks in the display to notify of the alarm (Figure
12.3).
Pressing the alarm indication displays a description of the alarm. Alarms include those shown in
Table 12.2.
Displayed alternately
AL-06
21.0%
F12.4E.ai
Figure 12.4
Table 12.2
Types of Alarms and Reasons for Occurrence
Alarm
Type of alarm
Reason for occurrence
Alarm 1
through 3
Oxygen concentration alarm
concentration alarm
humidity, and mixing ratio alarms
Occurs when a measured value exceed or falls below the
set alarm value (refer to Section 8.3, "Setting Alarms").
Alarm 6
Zero calibration coefficient alarm
Generated when the zero correction ratio is out of the
range of 100 ± 30% in automatic and semi-automatic
calibration (refer to Section 9.1.4, Compensation).
Alarm 7
Span calibration coefficient alarm
Generated when the span correction ratio is out of the
range of 0 ± 18% in automatic and semi-automatic
calibration (refer to Section 9.1.4, "Compensation").
Alarm 8
EMF stabilization time-up
Generated when the cell (sensor) voltage is not stabilized
even after the calibration time is up in automatic and
semi-automatic calibration.
Alarm 10 Cold junction temperature alarm
Occurs when an equipment internal temperature
exceeds 85°C.
Alarm 11
Thermocouple voltage alarm
Generated when thermocouple voltage exceeds 42.1 mV
(about 1020°C ) or falls below -5 mV (about -170°C).
Alarm 13 Battery low alarm
Internal battery needs replacement
If an alarm is generated, such measures as turning off the heater power are not carried out. The alarm
is released when the cause for the alarm is eliminated.
However, Alarm 10 and/or Alarm 11 may be generated at the same time as Error-2 (heater temperature
error).
In such a case, the measure taken for this error has priority.
If the converter power is turned off after an alarm is generated and restarted before the cause of the
alarm has been eliminated, the alarm will be generated again.
However, Alarms 6, 7, and 8 (alarms related to calibration) are not generated unless calibration is
executed.
12.2.2
Measures Taken When Alarms are Generated
l
Alarm 1 through 3: Oxygen Concentration Alarm, Humidity, and Mixing Ratio
Alarms
This alarm is generated when a measured value exceeds an alarm set point or falls below it. For
details on the oxygen concentration alarm, see Section 8.3, “Setting Alarms,” in the chapter on
operation.
Содержание EXAxt ZR402G
Страница 9: ...Blank Page...
Страница 51: ...Blank Page...
Страница 151: ...Blank Page...
Страница 163: ...CMPL 11M12A01 04E 9th Edition Feb 2016 YK Hood for ZR202G ZR202G_F ai 1 Item Part No Qty Description 1 K9472UF 1 Hood...
Страница 165: ......
Страница 167: ......