<10. Other Functions>
10-9
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
(1) Error 1
if the cell voltage is defective
(2) Error 2
if a temperature alarm occurs
(3) Error 3
if the A/D converter is defective
(4) Error 4
if an EEPROM write error occurs
For details on error occurrence, consult Chapter 12, “Troubleshooting”, later in this manual.
If any of the above problems occurs, the equipment turns off the power to the detector heater.
To cancel the error, reset the equipment following the steps below, or turn the power off and then
back on.
CAUTION
• Make sure that before resetting or restarting the power that there is no problem with the
equipment.
• If a problem arises again after the resetting, turn the power off and troubleshoot the problem
by consulting the Troubleshooting chapter later in this manual. When there is no error, the
Basic panel display will appear.
Table 10.6
Resetting
Switch operation
Display
Brief Description
>
∧
ENT
Err-01
/------
If an error occurs, the error number and "------" are displayed
alternately, as given on the left.
>
∧
ENT
PASSno
Hold down the [ENT] key for at least three seconds.
>
∧
ENT
0
000
Touch the [ENT] key again to switch to the password entry display.
>
∧
ENT
1
000
Enter the password 1102.
Intermediate switch operations omitted.
>
∧
ENT
1102
>
∧
ENT
A
01
>
∧
ENT
G
01
Change the parameter code to "G30".
>
∧
ENT
G
0
1
>
∧
ENT
G
3
0
>
∧
ENT
All the digits light
up.
Touch the [ENT] key to execute resetting.
The symbol [
] indicates that the corresponding keys are being touched, and the light characters indicate “ flashing.”
“ / ” indicates that the characters are displayed alternately.
CAUTION
Parameters of blank item are not used for Oxygen Analyzer or Humidity Analyzer.