<2. Specifications>
2-11
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
ZH21B.ai
t
C
C
øB
øB
øA
øB
D
ø72
ø76.3
428 (Insertion length)
Install facing upwards
JIS flange
ANSI flange
Unit: mm
Flange
A
B
C
t
D
JIS 5K 80 FF
180
145
4 - Ø19
12
40
ANSI Class 150 4B FF
228.6
190.5
8 - Ø19
12
50
2.2
ZA8F Flow Setting Unit and ZR20H
Automatic Calibration Unit
2.2.1
ZA8F Flow Setting Unit
This flow setting unit is applied to the reference gas and the calibration gas in a system
configuration (System 2). Used when instrument air is provided.
This unit consists of a flowmeter and flow control valves to control the flow of calibration gas and
reference gas.
Standard Specifications
FIowmeter Scale:
Calibration gas; 0.1 to 1.0 l/min. Reference gas; 0.1 to 1.0 l/min.
Construction:
Dust-proof and rainproof construction
Case Material:
SPCC (Cold rolled steel sheet)
Painting:
Baked epoxy resin, Dark-green (Munsell 2.0 GY 3.1/0.5 or equivalent)
Tube Connections:
Rc1/4 or 1/4NPT (Female)
Reference Gas Pressure: Clean air supply of sample gas pressure plus approx. 50 kPa G (or
sample gas pressure plus approx.150 kPa when a check valve is used.)
Pressure at inlet of the fl ow setting unit. (Max. 300 kPa G)
Air Consumption:
Approx. 1.5 l/min
Weight:
Approx. 2.3 kg
Calibration gas (zero gas, span gas) Consumption: Approx. 0.7 l/min (at calibration time only)
NOTE
Use instrument air for span calibration gas, if no instrument air is available, contact YOKOGAWA.
Содержание EXAxt ZR402G
Страница 9: ...Blank Page...
Страница 51: ...Blank Page...
Страница 151: ...Blank Page...
Страница 163: ...CMPL 11M12A01 04E 9th Edition Feb 2016 YK Hood for ZR202G ZR202G_F ai 1 Item Part No Qty Description 1 K9472UF 1 Hood...
Страница 165: ......
Страница 167: ......