<8. Detailed Data Setting>
8-6
IM 11M12A01-04E
11th Edition : Jul. 19, 2017-00
8.2.1
Definition of Equipment Status
(1) During warm-up
“During warm-up” is the time required after applying power until the sensor temperature
stabilizes at 750°C, and the equipment is in the measurement mode. This status is that the
sensor temperature is displayed on the basic panel.
(2) During maintenance
“During maintenance” is the time from when a valid password is entered in the basic panel
display to enable the parameter code selection display until the display goes back to the
basic panel display
(3) During calibration (see Chapter 9, Calibration)
In the manual calibration, proceed with the calibration operation with the parameter code
「
B10
」
to display the span gas confirmation display for the first span calibration, thus
starting the calibration time when the [ENT] key is touched. After a series of calibrations is
complete and the preset output stabilization time has elapsed, the calibration time will be up.
Figure 8.1 shows the definition of “during calibration” in the manual calibration.
Switch operation
Display
>
∧
ENT
b
1
0
>
∧
ENT
CAL
>
∧
ENT
CAL
>
∧
ENT
SPAn
Y
>
∧
ENT
21.00
%
>
∧
ENT
OPEn/20.84
>
∧
ENT
20.84
%
>
∧
ENT
ZEro
Y
>
∧
ENT
0.98
%
>
∧
ENT
OPEn/0.89
>
∧
ENT
0.89
%
>
∧
ENT
CALEnd
>
∧
ENT
b
10
>
∧
ENT
Measured-value
display
Figure 8.1
Definition of During Calibration
Output hold time
during calibration
Содержание EXAxt ZR402G
Страница 9: ...Blank Page...
Страница 51: ...Blank Page...
Страница 151: ...Blank Page...
Страница 163: ...CMPL 11M12A01 04E 9th Edition Feb 2016 YK Hood for ZR202G ZR202G_F ai 1 Item Part No Qty Description 1 K9472UF 1 Hood...
Страница 165: ......
Страница 167: ......