7
Precision Making
3. Sample rate calculation
ScopeCorder introduction workbook
Copyright © Yokogawa Electric Corporation
Sample Rate=
Main (Record Length)
Total Time
=
10.000 points
200
μ
s/DIV×10 DIV
=
10.000
2000
μ
s
=
10.000
0,002s
=5MS/s
What sample rate & measurement time
settings are possible?
Dependent on channel count, available
memory and other parameters different
sample rates & measurement times are
available. Review
information.