4
Precision Making
1. Factory reset
ScopeCorder introduction workbook
Copyright © Yokogawa Electric Corporation
A Factory RESET may be performed if the set-up status of the ScopeCorder is unknown.
It may be useful before performing these exercises, or before beginning a new test procedure.
Follow these steps:
1.
Turn OFF the ScopeCorder
2.
Hold the RESET button while turning ON the ScopeCorder
3.
Continue to hold the RESET button
4.
Wait for the confirmation message (you will hear a BEEP):
‘Turned on pressing the RESET key. Will initialize’
5.
Release the RESET button
6.
Press the ESC button
DONE! Factory reset has been completed.
Initialize vs. Factory Reset
A factory reset will initialize ALL
settings, including those of the
communication interfaces. To
maintain communication interface
settings, use
SETUP
»
INITIALIZE
.