SM 700820-01E
2 - 2
2.3 Equipment Required for Performance Testst
Equipment
Qty
Mandatory specifications
Recommended
DC voltage generator
1
Accuracy
±
0.05%
Yokogawa 7651
Output voltage –20 V to 20 V
Output resolution 1 mV
Standard signal generator
1
Output frequency range
HP 8657A
0.1 MHz to 150 MHz
Power meter
1
100 kHz to 150 MHz
HP 437B
Power sensor
1
100 kHz to 150 MHz
HP8482A
Function generator
1
Triangular wave, 100 kHz, 800 mV Yokogawa 7060
50
Ω
terminator
1
HP,1251-2277
Conversion adaptor
1
Banana plug to BNC
Yokogawa 3669 22
N-BNC conversion adaptor
1
N plug to BNC jack
HP 1250-0780
BNC-N conversion adaptor
1
N jack to BNC jack
HP 1250-1474
Cable
1
BNC to BNC
Yokogawa 3669 24
Probe
1
150 MHz frequency range
Yokogawa 700998
Logic Probe
4
8 bit
Yokogawa 700985
2.4 Self diagnosis
Equipment required
None
Procedure
Follow the procedure described in the Instruction Manual (IM 700820-01E) Section
14.3, “Self-diagnostics Test (self-test).”
2.5 Vertical Axis DC Voltage Accuracy Test
Specifications
2 mV/div:
±
(3% of 8 div. + 0.2 mV)
5 mV/div to 50 mV/div:
±
(1.5% of 8 div. + 0.2 mV)
100 mV/div to 500 mV/div:
±
(1.5% of 8 div. + 2 mV)
1 V/div to 5 V/div:
±
(1.5% of 8 div. + 20 mV)
Permissible range
Range
Tolerance
2 mV/div
±
0.68 mV
5 mV/div
±
0.8 mV
10 mV/div
±
1.4 mV
20 mV/div
±
2.6 mV
50 mV/div
±
6.2 mV
100 mV/div
±
14 mV
200 mV/div
±
26 mV
500 mV/div
±
62 mV
1 V/div
±
140 mV
2 V/div
±
260 mV
5 V/div
±
620 mV
Equipment required
Equipment
Qty
Mandatory specifications
Recommended
DC voltage generator
1
Accuracy
±
0.05%
Yokogawa 7651
Output voltage –20 V to 20 V
Output resolution 1 mV
Cable
1
BNC to BNC
Yokogawa 3669 24
Conversion adaptor
1
Banana plug to BNC
Yokogawa 3669 22
Содержание DL2700 Series