PERFORMANCE TEST
2 - 7
SM 700820-01E
2
2) Set the DL2700 as shown below.
VERTICAL
CH1
V/DIV
100 mV/div
COUPLING
DC
PROBE
1:1
HORIZONTAL
TIME/DIV
10
µ
s/div
RECORD LENGTH
10 K
DELAY
0
µ
s
TRIGGER
SIMPLE
MODE
NORMAL
SOURCE
CH1
MEASURE
MODE
ON
TRACE
CH1
ITEM
Select FREQ
3) Input the following signal to the DL2700 oscilloscope.
Frequency: 100.05 MHz
Amplitude: 0 dBm
The measured frequency of the beat wavefrom on screen by using MEASURE must be
50 kHz
±
10 kHz.
(
±
10 kHz tolerance is equal to
±
0.01% of the sampling clock frequency, 100MHz)
2.8 Trigger Sensitivity Test
Specifications
2 mV/div
DC to 150 MHz:
2 divp-p on the screen
Other ranges DC to 150 MHz:
1 divp-p on the screen
Permissible range
50 mV/div
150 MHz:
1 divp-p on the screen
Equipment required
Equipment
Qty
Mandatory specifications
Recommended
Standard signal generator
1
Output frequency range
HP 8657A
150 MHz
50
Ω
terminator
1
HP 1251-2277
N-BNC conversion adaptor
1
N plug to BNC jack
HP 1250-0780
Cable
1
BNC to BNC
Yokogawa 3669 24
2.7 Time -base Accuracy Test