PERFORMANCE TEST
2 - 9
SM 700820-01E
2
Permissible range(when 100 mV/div)
Trigger level
Permissible range
300 mV
170 mV to 430 mV
0 mV
–100 mV to 100 mV
–300 mV
–430 mV to –170 mV
Equipment required
Equipment
Qty
Mandatory specifications
Recommended
Function generator
1
Triangular wave, 100 kHz, 800 mV Yokogawa 7060
50
Ω
terminator
1
HP 1251-2277
Cable
1
BNC to BNC
Yokogawa 3669 24
Connection
50 ohm Termination
Function Generator
Cable
DL2700
Procedure
1) Turn on the power source of the DL2700. After warm-up press the CAL key, then press
the “Calibration” soft key to calibrate the instrument.
Furthermore, press the INITIALIZE key, then press the “Initialize” soft key to initialize
the settings.
2) Set the DL2700 oscilloscope as follows.
HORIZONTAL
TIME/DIV
5
µ
s
VERTICAL
CH1
V/DIV
100 mV/div
OFFSET
0.0V
COUPLING
DC
PROBE
1:1
TRIGGER
SIMPLE
LEVEL
300 mV, 0 V, –300 mV
MODE
NORMAL
SOUCE
CH1
SLOPE
,
CURSOR
TYPE
MARKER
CURSOR TRACE1
CH1
CURSOR1
0.0 s
2.9 Trigger Accuracy Test