SM 700820-01E
2 - 4
3) Input the following voltages from the voltage generator to the DL2700 to be tested, read
the indication on the DL2700 (value of Avg), and compare the reading with the toler-
ance.
Measurement range
Test input voltage
Tolerance
2m V/div
–6, 0, +6 mV
±
0.68 mV
5m V/div
–15, 0, +15 mV
±
0.8 mV
10 mV/div
–30, 0, +30 mV
±
1.4 mV
20 mV/div
–60, 0, +60 mV
±
2.6 mV
50 mV/div
–150, 0, +150 mV
±
6.2 mV
100 mV/div
–300, 0, +300 mV
±
14 mV
200 mV/div
–600, 0, +600 mV
±
26 mV
500 mV/div
–1500, 0, +1500 mV
±
62 mV
1 V/div
–3, 0, +3 V
±
140 mV
2 V/div
–5, 0, +5 V
±
260 mV
5 V/div
–15, 0, +15 V
±
620 mV
4) Test all channels in the same manner.
2.6 Frequency Response Test
Specifications
In equivalent sampling mode:
DC to 150 MHz (–3 dB point)
Permissible range
Input frequency
Permissible range
0.1 MHz
±
1 dB
10.02 MHz
±
1 dB
150.02 MHz
+1, –3 dB
Equipment required
Equipment
Qty
Mandatory specifications
Recommended
Standard signal generator
1
Output frequency range
HP 8657A
0.1 MHz to 150 MHz
Power meter
1
100 kHz to 150.02 MHz
HP 437B
Power sensor
1
100 kHz to 150.02 MHz
HP8482A
50
Ω
terminator
1
HP 1251-2277
N-BNC conversion adaptor
1
N plug to BNC jack
HP 1250-0780
BNC-N conversion adaptor
1
N jack to BNC jack
HP 1250-1474
Cable
1
BNC to BNC
Yokogawa 3669 24
Calibration of standard signal generator
Connection
Power Meter
Power Sensor
Standard Signal Generator
N-BNC
Conversion Adapter
BNC-N
Conversion Adapter