3 - 2
SM 700820-01E
3.3 Equipment Required
Table 3.1 Equipment required
Equipment
Critical Specification
Recommended
DC Volt Standard
Output Level : >1V
YOKOGAWA 7651
Accuracy : <0.02%
Digital Volt Meter
Accuracy : <0.01%
YOKOGAWA 7551
Function Generator
Frequency : 10 kHz
YOKOGAWA FG120
Output Level : >12 Vp-p
Frequency Counter
Frequency : >100 MHz
YOKOGAWA TC110
Accuracy : < 0.005%
BNC-BNC Cable
YOKOGAWA 3669 24
Clip-BNC Cable
YOKOGAWA 3669 26
50 ohm Terminators
Feed Thru Type
YOKOGAWA 7009 76
NOTE
The values shown in the specification column are those set in this service manual.
These values do not indicate the performances of the recommended equipment and
tools. Therefore, non-designated equipment and tools, which satisfy the specifications,
may be permitted for use.
The Function Generator setting values described in this service manual are those output
when 50 ohm load resistance is applied.
3.4 Dip Switch Setting on the CPU Board
Procedure
1) Remove the main unit cover and the two fixing plates.
2) Set the switch on the CPU Board (B9929LN/B9929LP) as following condition.
bit 1: OFF
(not used)
bit 2: OFF
(not used)
bit 3: OFF
(not used)
bit 4: ON
English Help Message
OFF
Japanese Help Message
3.5 Rotary Switch Setting on the Printer Board
Procedure
1) Remove the main unit cover and the two fixing plates.
2) Set the switch on the Printer Board (B9919LF) corresponding to the Head Rank of the
attached Printer unit.
Rank A:
A
Rank B:
B
Rank C:
C
Rank D: D
Rank E:
E
Rank F:
F
Rank G: 6
Rank H: 7