5-28
IM CW500-01EN
Record Method Settings
Interval
Set the interval to record the measured data to the SD memory card
or internal memory. Seventeen different intervals are available, but if
the measurement interval of demand measurement is already set, the
record interval cannot be set longer than the measurement interval.
To set a longer interval, change the measurement interval of demand
measurement to “Not be used” first. If you set the measurement interval of
demand measurement after you set the record interval, the record interval
is automatically set to the same setting as the measurement interval. For
details, see “Demand Measurement Settings” (page 5-12).
Settings
1s/2s/5s/10s/15s/20s/30s/
1min/2min/5min/10min/15min/20min/30min*/
1h/2h/150, 180 cycle (approx. 3 s)
* Default value
The intervals 150, 180 cycles (approx. 3 sec) are defined in the
international power quality standard IEC61000-4-30. Data will be
collected for 150 cycles for 50 Hz nominal frequency and 180 cycles for
60 Hz nominal frequency.
<Procedure>
Use the arrow keys to move to Interval.
→
Press ENTER to show a
selection window.*
→
Use the arrow keys to enter the recording interval.
→
Press ENTER to confirm or ESC to cancel.
5.4 Recording Setup