14-8
IM CW120-E
14.2 Specifications of Clamp-on Current
Probes
(1) 96030 and 96031 Clamp-on Current Probes
Amplitude
Phase
Item
96030 (Max. 200-A Range)
96031 (Max. 500-A Range)
Measurement range
Output voltage
Accuracy
Temperature coefficient
Maximum allowed current
Output impedance
External magnetic field effects
Conductor position effects
Circuit voltage
Withstanding voltage
Measurable conductor
diameter
Operating temperature
and humidity ranges
Storage temperature and
humidity ranges
Applicable locations
Dimensions (W
×
H
×
D)
Weight
Output terminals
Output cable length
Standard accessories
0 to 200 Arms AC (300 Apk)
0 to 0.5 Vrms AC (2.5 mV/A)
±
1.5% of rdg
±
0.4 mV (20 Hz to 45 Hz)
±
0.5% of rdg
±
0.1 mV (45 Hz to 66 Hz)
±
0.8% of rdg
±
0.2 mV (66 Hz to 1 kHz)
±
2.0% of rdg
±
0.4 mV (1 kHz to 20 kHz)
±
0.5
°
(45 Hz to 1 kHz)
Under the conditions of the temperature at 23
°
C
±
5
°
C, humidity at 35 to 75% RH, with sine
wave input
0.05% of full scale/
°
C in ranges of 5
°
C to 40
°
C
250 Arms AC (45 Hz to 1 kHz)
Approximately 6
Ω
0.1 A equivalent or less (at 400 A/m, 50/60 Hz)
±
0.5% (at 20 to 200 A, 45 Hz to 1 kHz)
600 Vrms AC maximum
3.7 kVrms AC for 1 minute (core-to-casing, and core-to-output)
30 mm diameter maximum
5
°
to 40
°
C, 35 to 80% RH (no condensation)
–20
°
to 60
°
C, 90% RH or less (no condensation)
Indoors, altitude of 2000 m or lower
Approximately 73
×
130
×
30 mm
Approximately 300 g
Banana plugs (safety terminals)
Approximately 3 meters
One User’s Manual, ring markers (4 colors
×
2 each, L4007MG)
0 to 500 Arms AC (750 Apk)
0 to 0.5 Vrms AC (1 mV/A)
±
1.5% of rdg
±
0.4 mV (20 Hz to 45 Hz)
±
0.5% of rdg
±
0.1 mV (45 Hz to 66 Hz)
±
0.8% of rdg
±
0.2 mV (66 Hz to 1 kHz)
±
1
°
(45 Hz to 1 kHz)
625 Arms AC (45 to 400 Hz)
Approximately 2.4
Ω
0.2 A equivalent or less (at 400 A/m, 50/60 Hz)
±
0.5% (at 50 to 500 A, 45 Hz to 1 kHz)
Safety standards:
Compliant with EN61010-1, EN61010-2-032
• Overvoltage category II (CAT II, max. input
voltage 600 Vrms)
• Overvoltage category III (CAT III, max. input
voltage 300 Vrms)
• Pollution degree 2
EMC (emission):
Compliant with EN55011, Group 1, Class A; and
EN61326: 1997
These products are goods in Class A (designed
for industrial environments). For home use, the
user must implement adequate measures to
prevent influences from radio noise emission.