14-1
IM CW120-E
Specifications
14
Chapter 14. Specifications
14.1 Specifications of CW120/121
(1) Inputs
Item
Voltage Inputs
Current Inputs
Input type
Ratings (ranges)
Phase lines
Input
resistance
Maximum allowed
continuous input
A/D conversion
CW120
CW121
CW120
CW121
Resistive potential division
150, 300, 450 V
Single-phase two wires, single-phase three wires, three-phase three wires
Single-phase two wires, single-phase three wires, three-phase three wires,
three-phase four wires
Approximately 1.5 M
Ω
Approximately 1.3 M
Ω
495 Vrms
Simultaneous voltage/current input conversion, 12-bit resolution
Clamp sensing
96033 clamp-on probe: 5, 10, 20, 50 A
96030 clamp-on probe: 20, 50, 100, 200 A
96031 clamp-on probe: 50, 100, 200, 500 A
96032 clamp-on probe: 200, 500, 1000 A
96033 clamp-on probe: 130 Arms
96030 clamp-on probe: 250 Arms
96031 clamp-on probe: 625 Arms
96032 clamp-on probe: 1000 Arms
(2) Measurement Functions
Item
Voltage
Current/Active Power
Method
Frequency range
Crest factor
Effective input range
Display range
Temperature coefficient
Display update period
Digital sampling
45 to 65 Hz (reciprocal method), detected from V1 input
150 or 300 V range: 2 for rated input
450 V range: 1.56 for rated input
10% to 110% of each rated range
1.5 V for all rated ranges
130% of each rated range for all ranges
other than 450 V
110% of rated range for 450-V range
±
0.05% of rng (range)/
°
C
Approximately 1 second
3 for rated input
0.4% of each range
130% of each rated range
±
0.07% of rng (range)/
°
C
(including clamp-on probe)
Lower limit
Upper limit