<3.Construction and Function>
3-2
IM 11R02A01-02E
5th Edition : Dec. 28, 2016-00
l
Town Gas Use
The sample gas, introduced through line filter (3-1), increases its pressure by pump, or
decreases by pressure reducing valves, according to the pressure at the sampling point. The
pressure gauge (3-2) indicates 8 to 18 kPa and the flowmeter (3-3) approx. 10 l/min, respectively.
The sample gas, then, is set its differential pressure to 500 Pa by pressure reducing valve (3-4)
and (3-6), and at this time the pressure gauge (3-5) indicates approx. 3 kPa. When the density
compensation system is equipped, the sample gas is introduced to the density meter with the flow
rate of 1 l/min, through the flowmeter (5-3).
The standard gas is reduced its pressure to 8 to 18 kPa by the pressure reducing valve (4-1), and
supplied, same as the sample gas, with the flow rate approx. 10 l/min.
l
Steel Mill Use
The pressure of the sample gas is increased, by the pump, according to that of the sampling
point.
The pressure gauge (3-1) indicates approx. 6 kPa. The sample gas then flows through the
washing bubbler (3-2) and the fulflo filter (3-3), and then secure a constant pressure in the
pressure regulating pot (3-4) through the water sealed pipe from the dehumidifier (3-5), then set
the differential pressure by the pressure reducing valve (3-6) to 500 Pa.
To increase the pressure by the pump, a drain pot (3-9) is added. When the density
compensation system is equipped, the sample gas is supplied to the density meter with its flow
rate of 1 l/min, through the flowmeter (5-3). The pressure of the standard gas is reduced by the
pressure reducing valve (4-1) to about 6 kPa and is supplied by the flowmeter (4-2) with the flow
rate about 10 l/min.
Standard Flow Sheet
1. Town Gas Application (Standard Type)
V-3
V-2
V-1
V-4
3-1
V-5
V-12
1-6
7-2
7-1
6-2
6-1
4-1
INST.AIR IN
0.3~0.7MPa
50Nl/min
SAMPLE GAS IN
STD GAS IN (ZERO)
SAMPLE GAS OUT
OR VENT
V-6
V-11
V-7
V-13
V-14
V-9
V-10
V-8
V-16
V-15
5-3
3-4
3-6
3-3
3-2
3-5
1-4
1-5
GAS
DP.T
AIR
DP.T
1-3
1-2
1-1
2-4
2-3
2-1
2-5
2-2
3-7
No. Item
1-1 Calorie detector
1-2 Solenoid valve
1-3 Orifice assembly
1-4 Differential pressure transmitter (air)
1-5 Differential pressure transmitter (gas)
1-6 Computing station
2-1 Air set
2-2 Pressure reducing valve
2-3 Pressure gauge
2-4 Preheating chamber
2-5 One touch coupler
No. Item
3-1 Line filter
3-2 Pressure gauge
3-3 Flowmeter
3-4 Pressure reducing valve
3-5 Pressure gauge
3-6 Pressure reducing valve
3-7 Diaphragm pump (when specified)
3-8 Pressure reducing valve (when specified)
4-1 Pressure reducing valve for standard gas
No. Item
5-1 Density meter detector (when specified)
5-2 Density meter converter (when specified)
5-3 Flowmeter for density meter (when specified)
6-1 Zero gas (supplied by customer)
6-2 Span gas (supplied by customer)
7-1 Pressure reducing valve for cylinder (optional)
7-2 Pressure reducing valve for cylinder (optional)
V1...16 Ball valve, needle valve
*1
*2
3-8
*3
*1: CM6G-S6
10 (with density meter)
*2: CM6G-S63
0 (with diaphragm pump)
*3: CM6G-S64
0 (with pressure reducing valve)
CM6G-S6200, S6210, S6300, S6310, S6400, S6410
A
B
C
D
E
DT
P
5-2
5-1
Maximum back pressure
is 1.5 kPa and no pressure
fluctuation is allowed.
FLOW-6410.ai
STD GAS IN (SPAN)
Figure 3.2 Flow Sheet (for Town Gas)