<7. Maintenance>
7-3
IM 11R02A01-02E
5th Edition : Dec. 28, 2016-00
NOTE
The orifice system is temperature-controlled by a plate heater, so better finish the cleaning and
replacement as swiftly as possible.
7.2.2
Fulflo Filter (For Steel Mill Use)
The material of the element is polypropylene. The cleaning and the replacement of the element
shall be carried out as follows. Same process applies to the option fulflo filter whose element
(50µm) is G7057XJ in the part number.
1) Turn the nut and remove the cover.
2) Pull out the element and clean or replace.
3) Assembly is carried out in contrary order to above.
4) When the packing or gasket is worn out, replace it. We provide a packet which contains 4
pieces of packing and a gasket. (G7086XL)
F0703.ai
PACKING GASKET
ELEMENT
PACKING
NUT
COVER
CASE
TENTION BOLT
Figure 7.3 Fulflo Filter
7.2.3
Line Filter (For Town Gas Use)
Disassemble 3 screws, and open the cover, there inside exist an element (G7005XJ.) If it is
stained, replace by a new one. Also if the O-ring (Y9116XB) is fatigued, replace it with new one.
SCREW
FILTER
ELEMENT
O-RING
GAS INLET
BODY
GAS OUTLET
ELEMENT
GUIDE
F11-5.ai
Figure 7.4 Line Filter