2-3
IM AQ6360-02EN
Preparing for Measurement
1
2
3
4
5
App
Index
Also be sure to maintain sufficient clearance for connecting measurement cables.
20 cm or
more
20 cm or
more
Inlet holes (also on the bottom panel)
20 cm or more
20 cm or
more
Ambient Temperature and Humidity
Ambient temperature: 5 to 35°C
Ambient humidity: 20%RH to 85%RH (no condensation present)
Note
Condensation may occur if the instrument is moved to another place where the ambient
temperature is higher, or if the temperature changes rapidly. In such cases, allow sufficient time
for the instrument to adjust to the ambient temperature before use.
When the instrument is packaged in a box and moved, prevent condensation by allowing
sufficient time for the instrument to acclimatize before removing it from the box.
Flat, Even Surface
Install the instrument with the correct orientation on a stable, horizontal surface. If the
instrument is installed in a horizontal position, rubber stoppers can be attached to the
feet at the rear of the instrument to prevent the instrument from sliding. One set of rubber
stoppers (two stoppers) are included with the instrument.
Rubber stoppers
A9088ZM
Feet at the rear of the instrument
Rack Mounting
To rack-mount the instrument, use the separately sold rack mount kit.
Product Name
Model
Rack Mounting Kit for EIA Single
751535-E4
Rack Mounting Kit for JIS Single
751535-J4
2.1 Installing the Instrument