2.2 Attaching the Connector Adapter
Attach the optional connector adapter before using the instrument.
WARNING
Always turn the power OFF before replacing the connector adapter. This instrument
has a built-in reference light source for wavelength calibration (/LFC, /LSC option)
and transmits infrared light from the optical output connector at all times. Never
look into the optical output connector. Infrared light entering the eyes can cause
severe injury and loss of vision.
CAUTION
• As there may be dust adhering to calibration output, be sure to clean it before
attaching the connector adapter.
• Do not exhale or blow compressed air into the monochromator from the optical
input. Doing so may allow dust or other materials to enter the monochromator,
adversely affecting its optical performance. Also, if debris is adhering to the
optical components inside the monochromator when a strong light source is
input, the monochromator may be damaged.
• When attaching or removing the connector adapter, carefully insert it
perpendicularly to the ferrule so as not to damage the ferrule end.
• Moving the connector adapter to the right or left or inserting it forcefully can
damage the ferrule or the connector adapter.
French
AVERTISSEMENT
Toujours éteindre l’avant de remplacer l’adaptateur de connecteur.
Cet instrument dispose d’une source de lumière de référence intégrée pour les
ajustements d’alignement (/LFC, /LSC option). La lumière infrarouge est toujours
émise depuis le connecteur de sortie optique. Ne regardez jamais directement
dans le connecteur de sortie optique. La lumière infrarouge risquerait de gravement
vous blesser ou de provoquer une perte de vision.