2-13
IM AQ6360-02EN
Preparing for Measurement
1
2
3
4
5
App
Index
Operations Performed When the power is Turned On
When the initialization finishes successfully, the following measurement screen appears.
Appears only on models
with an /LFC or /LSC
option (with built-in
reference light source).
On models with an /LFC or /LSC option (with built-in reference light source), a message
prompting you to execute wavelength calibration appears. The message that appears is
shown below.
For this instrument to meet its specifications, a Wavelength Calibration must be
performed. Please perform according to the guidelines below.
Wavelength
Calibration
Please perform a Wavelength Calibration after a 1-hour warm-up and before
starting a measurement. Unless the Wavelength Calibration is carried out, the
wavelength accuracy of this instrument cannot be guaranteed.
For details on wavelength calibration, see section 2.1 in IM AQ6360-01EN.
When the Power-on Operation Does Not Finish Normally
Turn off the power switch, and check that :
• The instrument is installed properly. See section 2.1, "Installing the Instrument."
• The power cord is connected properly. See the previous page.
If the instrument still does not work properly, contact your nearest YOKOGAWA dealer for
repairs.
If a memory error or other errors occur during initialization, initialization is cancelled at
the step in which the error occurs (STEP @/9, where @ is a number from 1 to 9). Then,
the screen changes to the measurement screen, and a warning message is displayed.
If this occurs, repairs are necessary. Contact your nearest YOKOGAWA dealer
immediately.
Note
The instrument retains the measurement conditions, the function menu state, the waveforms
that it is showing, and so on. When the power is turned ON, the state of the instrument prior to
the last shut down is restored. When the power is turned ON for the first time, the instrument
starts up in the factory default state.
2.4 Turning the Power ON/OFF