10-62
IM 701730-01E
• Exiting the Zone Editing
18.
Press the
Quit & To TopMenu
soft key. The screen returns to the menu shown in
step 3.
If you press the Quit & To TopMenu soft key without registering the edited zone in
steps 15 to 17, the editing up to that point is discarded, and the screen returns to
the menu of step 3.
• Modifying the Registered Zone
Turn on the display of the reference waveform that was selected when the zone was
created, and then correct the zone.
19.
At step 5, press the
Redraw
soft key. A menu for selecting the registered zone
appears.
20.
Repeat steps 7 and 18 to modify the registered zone.
The corrected zone can be registered to a different destination.
Setting the Determination Conditions
21.
At step 4, press the
Setup
soft key. The Setup dialog box opens.
22.
Use
jog shuttle & SELECT
to set the following items of determination zone Zone1.
•
Select the determination target waveform from CH1 to CH4 (2), Math1, or Math2 (Trace box).
•
Select whether to perform determination at the corresponding determination zone (ON/
OFF) (Mode box).
•
Select whether the waveform exiting the determination zone (OUT) or entering the
determination zone (IN) (Condition box) makes the condition true.
*
On the DL1735E/DL1740E/DL1740EL, you can select from CH1 to CH4, Math1, and
Math2. On the DL1720E, you can select CH1, CH2, and Math1.
23.
Repeat step 22 to set determination zones Zone2 to Zone6.
24.
Use
jog shuttle & SELECT
to set the determination logic to AND or OR (Logic
box).
25.
Use
jog shuttle & SELECT
to set the determination operation to Single or
Continue (Sequence box).
26.
Use
jog shuttle & SELECT
to set the waveform acquisition count (ACQ Count
box).
27.
Use
jog shuttle & SELECT
set the instrument action when the condition is met
(Action box).
Select the determination target
waveform from CH1 to CH4 (2),
Math1, or Math2.
Select which condition when the
waveform exits the determination
zone (OUT) or when the waveform
enters the determination zone (IN) is
considered true.
Select whether to perform determination at the
corresponding determination zone (ON/OFF).
Determination
zone Zone1
Determination
zone Zone2
Determination
zone Zone3
Determination
zone Zone4
Determination
zone Zone5
Determination
zone Zone6
Set the determination logic to AND or
OR.
Set the determination operation to
Single or Continue.
Set the number of waveform
acquisitions.
Set the action taken by the instrument
when the condition is met.
10.9 Performing GO/NO-GO Determination Using Zones