6-15
IM 701730-01E
Triggers
6
6.8 Setting the A->B(N) Trigger (ENHANCED)
Setting the Trigger Level
9.
Press the
Level/Coupling
soft key. The Level/Coupling dialog box opens.
The settings related to CH3 and CH4 are not displayed on the DL1720E.
*1. Setting the level of each channel
*2. Set the hysteresis of each
channel to or .
*3. Set the trigger coupling of each
channel to DC or AC.
*4. Set the HF rejection of each
channel to OFF, 20 MHz, or 15
kHz.
*1
*2
*3
*4
10.
Use
jog shuttle & SELECT
to set the level of each channel.
Setting the Hysteresis
11.
Use
jog shuttle & SELECT
to set the hysteresis of each channel to
or
.
Setting the Trigger Coupling
12.
Use
jog shuttle & SELECT
to set the trigger coupling of each channel to DC or
AC.
Setting the HF Rejection
13.
Use
jog shuttle & SELECT
to set the HF rejection of each channel to OFF,
20MHz, or 15kHz.
14.
Press
ESC
. The Level/Coupling dialog box closes.
Setting the Number of Times Condition B Is to Be Met
15.
Press the
Count
soft key.
16.
Turn the
jog shuttle
to set the count.
Pressing RESET resets the value to 1.
Setting the Hold Off
17.
Set the hold off time according to the procedures given in section 6.4.
Note
•
The status settings of conditions A and B apply to the A->B(N) trigger and A Delay B trigger.
•
The trigger level, hysteresis, trigger coupling, and HF rejection settings apply to both simple
and enhanced triggers.