4. Move the nib gradually to approach to the patient’s eyes: when the patient sees the nib becomes
two from one, stop moving the nib.
5. Measure the distance from the nib to the nose root of the patient.
6. Input the distance (in cm) into the input frame, the meter angle and prismatic power will be
calculated automatically.
5.5.15 Near –point adjustment (NPA) test
Test purpose: to test near-point adjustment
Test sighting mark: Letter sighting mark of short-distance visual chart.
Confirmation before test: Set “SPH distant vision
→
near vision” to SPH+ADD.
1. Press
,choose “Near Point”, NPC, NPA, NRA, PRA showing simultaneously, to
shift to short-distance mode.
2. Press “NPA” to open binocular testing window.
3. Place the visual chart in the distance of 40cm, make sure the patient see clearly the letter sighting
mark with best vision.
3. Move the visual chart gradually to approach to the patient’s eyes: when the sighting mark
becomes obscure, stop moving the visual chart.
4. Measure the distance from the visual chart to the nose root of the patient.
5. Input the distance (in cm) into the input frame, the near-point adjustment value will be calculated
automatically.
5.5.16 Negative-relative adjustment (NRA) test
Test purpose: to test the negative-relative adjustment when the two eyes congregate in specified
working distance
Test sighting mark: column sighting marks in short-distance visual chart
Содержание YPA-2100
Страница 1: ...YPA 2100 DIGITAL REFRACTOR User Manual Version 1 0 Date 20180521...
Страница 5: ...13 Electrical Schematic Diagram 82 14 Electromagnetic Compatibility 83...
Страница 24: ...19 Press to enlarge PD press to reduce PD 3 Press to exit PD input interface...
Страница 31: ...26 key...
Страница 42: ...37 4 Click key shift to 0 5XC 5 Click the key again shift to automatic mode Slit Prism...
Страница 55: ...50 34 Press to finish the test...
Страница 72: ...67 3 Program 1 A B C D are recommended vision test steps users can also add or delete as needed...
Страница 74: ...69 Add Key Click it to add one step automatically 4 Click key to enter program customizing interface...
Страница 78: ...73 3 Click key to popup input box...
Страница 79: ...74 4 Use CCQ 800 lensmeter to test lens data and convey the date to YPA 2100 5 Click key 6 Choose No00001...
Страница 80: ...75 7 Click key to complete data import...