33
8.Press to test the astigmatic power with astigmatism disk.
Take -0.25D cylinder as increment, gradually adjust the cylindrical power until the definitions of
lines to all directions in the astigmatism disc are the same.
9. Press message prompts: skip crisscross cylinder test?
When the cylindrical power is 0, message “skip crisscross cylinder test?” prompts.
Select “Yes”, skip to step 14. Select “No”, skip to step 10.