51
5.5 Binocular Functional Testing Method
5.5.1 Crisscross matrix test (presbyopia)
Test purpose: to test the spherical power.
Test sighting mark: crisscross matric sighting mark
Assistant lens: binocular ±0.50D fixed crisscross cylinder
1. Binocular distant vision test is finished, add distant vision degree in the testing window.
2. Add binocular ±0.50D fixed crisscross cylinder.
3. Lower down the short-distance vision rod and place the visual chart in necessary working
distance (generally 400mm).
4. Select the crisscross matrix sighting mark on the short-distance visual chart.
5. Ask the patient: the horizontal line or the vertical line, which is more distinct? Or the horizontal
line and the vertical line look the same?
If the horizontal line and the vertical line look the same: there is no need to carry out near vision test
and wear presbyopia glasses.
If the horizontal line is more distinct than the vertical line: increase two eyes by +0.25D additional
degree synchronically until the horizontal line looks as distinct as the vertical line.
6. Remove ±0.50D fixed crisscross cylinder.
7. Select short-distance visual chart to test the vision and slightly adjust the sphere to make the
patient see the sighting mark clearly to reach ideal vision value.
5.5.2 Cross test (heterophoria)
Test purpose: To test heterophoria
Test sighting mark: Cross sighting mark
Assistant lens: Binocular rotatory prime
Right eye red optical filter, left eye green optical filter (red and green cross sighting mark)
Binocular circular polarized optical filter (polarized cross sighting mark)
1. Call in cross sighting mark.
2. Ask the patient about the sighting mark he sees and test according his answers.
Содержание YPA-2100
Страница 1: ...YPA 2100 DIGITAL REFRACTOR User Manual Version 1 0 Date 20180521...
Страница 5: ...13 Electrical Schematic Diagram 82 14 Electromagnetic Compatibility 83...
Страница 24: ...19 Press to enlarge PD press to reduce PD 3 Press to exit PD input interface...
Страница 31: ...26 key...
Страница 42: ...37 4 Click key shift to 0 5XC 5 Click the key again shift to automatic mode Slit Prism...
Страница 55: ...50 34 Press to finish the test...
Страница 72: ...67 3 Program 1 A B C D are recommended vision test steps users can also add or delete as needed...
Страница 74: ...69 Add Key Click it to add one step automatically 4 Click key to enter program customizing interface...
Страница 78: ...73 3 Click key to popup input box...
Страница 79: ...74 4 Use CCQ 800 lensmeter to test lens data and convey the date to YPA 2100 5 Click key 6 Choose No00001...
Страница 80: ...75 7 Click key to complete data import...