36
12. Press
to make the crisscross cylinder accurately test the astigmatic axis of the right eye.
1
)
Call in speckle sighting mark and ±0.25D crisscross cylinder.
2
)
Reverse side 1 and side 2 of the crisscross cylinder, ask the patient: which side is more distinct?
When the side 1 is more distinct: enlarge the axis;
When the side 2 is more distinct: reduce the axis.
3
)
Repeat above steps until the definitions of side 1 and 2 look the same.