1. Specifications
1.1 Measurement range
The measuring range conforms to the requirements in Table 1.
Table 1 Vision tester measurement range
Item Measuring
range
Spherical power
-29.00D~+26.75D, Step size: 0.12D, 0.25D, 1D
Cylindrical power
-
8.75D
~+
8.75D, Step size: 0.25D, 1D
Cylindrical axis
0~180°, Step size: 1°, 5°
Prismatic power
0~20
△
, Step size: 0.1
△
, 0.5
△
Prism base
0~360°, Step size: 1°, 5°. Prim base can be marked horizontally or
vertically.
1.2 Host
1.2.1 Assistant lens
a. Automatic crisscross cylinder: automatic crisscross cylinder ±0.25D and ±0.50D, one at left, one
at right (the axle is self-rotatory);
b. Fixed crisscross cylinder: fixed crisscross cylinder ±0.50D, one at left, one at right (the axle is
fixed at 90°)
c. Pinhole plate: hole diameter 1mm, one at left, one at right;
d. ross plate: used when setting pupil distance, one at left, one at right;
e. Red and green optical filter: one is red optical filter (right) and one is green optical filter (left);
f. Inear polarized optical filter: one at 135° right and one at 45° left
One at 135° right/sphere 0.125D, one at 45° left/sphere 0.125D;
g.Circular polarized optical filter: two at left, two at right, one includes sphere 0.125D;
h. Maddox rod lens: right eye: red, horizontally, left eye: white, vertically;
i. Sphere for retinoscopy: +1.5D and +2.0D, one at left, one at right;
j. Decomposing prism: right eye: 6
△
BU, right eye: 10
△
BI
k. Single-slit lens: used for quick astigmatism test, one at left and one at right.
Содержание YPA-2100
Страница 1: ...YPA 2100 DIGITAL REFRACTOR User Manual Version 1 0 Date 20180521...
Страница 5: ...13 Electrical Schematic Diagram 82 14 Electromagnetic Compatibility 83...
Страница 24: ...19 Press to enlarge PD press to reduce PD 3 Press to exit PD input interface...
Страница 31: ...26 key...
Страница 42: ...37 4 Click key shift to 0 5XC 5 Click the key again shift to automatic mode Slit Prism...
Страница 55: ...50 34 Press to finish the test...
Страница 72: ...67 3 Program 1 A B C D are recommended vision test steps users can also add or delete as needed...
Страница 74: ...69 Add Key Click it to add one step automatically 4 Click key to enter program customizing interface...
Страница 78: ...73 3 Click key to popup input box...
Страница 79: ...74 4 Use CCQ 800 lensmeter to test lens data and convey the date to YPA 2100 5 Click key 6 Choose No00001...
Страница 80: ...75 7 Click key to complete data import...